Triglycerides

Catalog Number: 20767107322

SYSTEM INFORMATION

For cobas c 311/501 analyzers:

INTENDED USE

In vitro test for the quantitative determination of triglycerides in human serum and plasma on Roche/Hitachi cobas c systems.

STORAGE AND STABILITY

 TRIGL
Shelf life at 2‑8 °C: See expiration date on cobas c pack label.
On board in use and refrigerated on the analyzer: 8 weeks
Diluent NaCl 9 %
Shelf life at 2‑8 °C: See expiration date on cobas c pack label.
On‑board in use and refrigerated on the analyzer: 12 weeks

Specimen collection and preparation

APPLICATION FOR SERUM AND PLASMA

 cobas c 311 test definition
Assay type 1‑Point
Reaction time / Assay points 10 / 57
Wavelength (sub/main) 700 / 505 nm
Reaction direction Increase
Units mmol/L (mg/dL, g/L)
Reagent pipetting Diluent (H2O)
R1 120 µL 28 µL
Sample volumes Sample Sample dilution
Sample Diluent (NaCl)
Normal 2 µL
Decreased 4 µL 15 µL 135 µL
Increased 2 µL
 cobas c 501 test definition
Assay type 1‑Point
Reaction time / Assay points 10 / 70
Wavelength (sub/main) 700 / 505 nm
Reaction direction Increase
Units mmol/L (mg/dL, g/L)
Reagent pipetting Diluent (H2O)
R1 120 µL 28 µL
Sample volumes Sample Sample dilution
Sample Diluent (NaCl)
Normal 2 µL
Decreased 4 µL 15 µL 135 µL
Increased 2 µL

 

CALIBRATION

Calibration interval may be extended based on acceptable verification of calibration by the laboratory.

 Calibrators S1: H2OS2: C.f.a.s.
Calibration mode Linear
Calibration frequency 2‑point calibration- after reagent lot change- as required following quality control procedures

CALIBRATION MODE

CALIBRATION FREQUENCY

2‑POINT CALIBRATION

CALIBRATION INTERVAL MAY BE EXTENDED BASED ON ACCEPTABLE VERIFICATION OF CALIBRATION BY THE LABORATORY.

LIMITS AND RANGES

Measuring range

MEASURING RANGE

0.1‑10.0 mmol/L (8.85‑885 mg/dL)

LOWER LIMITS OF MEASUREMENT

Lower detection limit of the test

SPECIFIC PERFORMANCE DATA

Representative performance data on the analyzers are given below. Results obtained in individual laboratories may differ.

PRECISION

Precision was determined using human samples and controls in an internal protocol with repeatability (n = 21) and intermediate precision (3 aliquots per run, 1 run per day, 21 days). The following results were obtained:

PRECISION WAS DETERMINED USING HUMAN SAMPLES AND CONTROLS IN AN INTERNAL PROTOCOL WITH REPEATABILITY (N = 21) AND INTERMEDIATE PRECISION (3 ALIQUOTS PER RUN, 1 RUN PER DAY, 21 DAYS). THE FOLLOWING RESULTS WERE OBTAINED:

INTERMEDIATE PRECISION

METHOD COMPARISON

Triglycerides values for human serum and plasma samples obtained on a Roche/Hitachi cobas c 501 analyzer (y) were compared with those determined using the corresponding reagent on a Roche/Hitachi 917 analyzer (x).

BABLOK W, PASSING H, BENDER R, ET AL. A GENERAL REGRESSION PROCEDURE FOR METHOD TRANSFORMATION. APPLICATION OF LINEAR REGRESSION PROCEDURES FOR METHOD COMPARISON STUDIES IN CLINICAL CHEMISTRY, PART III. J CLIN CHEM CLIN BIOCHEM 1988 NOV;26(11):783-790.