td-systems-benchmark-ultra-hero.png

 

Contact a Rep HTML

Request more information for the BenchMark ULTRA system.

Please complete the form below to speak with a Roche representative.

 
 
 
 
 
 
Overview

True continuous-access automated IHC/ISH slide staining
for unprecedented workflow efficiency


Maximize Capacity

Individual slide drawers enable single piece flow to limit wasted slide positions.

Optimize Efficiency

Continuous access to each slide and protocol flexibility without sacrificing stain optimization.

Improve Processing Time

Decreased preparation time required for slide sorting and matching of reagent.

Quickly Respond to Test Requests

Ultimate Reagent Access (URA) allows immediate processing of urgent and late-arriving samples.

Increase Workflow Flexibility

During a run, users have continuous access to bulks and waste.

Explore

BenchMark ULTRA system

Unprecedented workflow

IHC, ISH, SISH, Dual Stain and FITC slides can be processed simultaneously, improving turnaround times. The fully automated workflow enables processing of 90 slides in 8 hours, or 120 slides with an overnight run.

Real-time remote monitoring and technical assistance

The BenchMark ULTRA system is equipped with CareGiver remote monitoring technology, offering 24/7 online support for basic troubleshooting and technical guidance. This provides proactive instrument service and maintenance, helping expedite recovery times to minimize system downtime.

Performance

BenchMark ULTRA Slide Staining System Technical Specifications

Unique features
  • Automated IHC/ISH platform manages STAT requests with no impact to in-process slides
  • Exceptional throughput — 90 slides per 8 hour shift
  • Individual slide heaters which eliminates need to batch runs

General characteristics

Fully automated Baking, deparaffinization, cell conditioning and staining, including IHC, ISH, SISH, immunofluorescence, multiparameter IHC staining (dual, triple), counter stain and titration
Slide position 1–30 slides with independent processing/functionality and temperature control for each position
Reagent carousel 35 reagent positions
Slides 25 x 75 mm, 1 x 3 in or 26 x 76 mm positively charged slides
Bulk solutions Up to 7 different bulk solutions can be changed “on the fly” without process interruption
Waste Two 20 L carboys, can be changed "on the fly" without process interruption
DI Water supply NCCLS Type II water or equivalent (deionized water)
Temperature control Ambient to 100°C
Modularity One to eight VENTANA BenchMark ULTRA, VENTANA Discovery ULTRA or VENTANA BenchMark Special Stains instruments may be controlled from one BenchMark ULTRA computer system
Configuration Free standing
Certifications EC/EN 61010-1, IEC/EN 61010-2-010, IEC/EN 61010-2-101, IEC/EN 61010-2-081, EN 61326-2-6
Catalog number 05342716001

Electrical specification

Voltage ~100-230V
Current 4–6 Amps
Frequency 50/60 Hz

Physical characteristics

Size (W x D x H) 44.00 in x 33.10 in x 62.40 in (111.76 cm x 84.07 cm x 158.50 cm)
13.5 in x 14.5 in x 13.5 in (34 cm x 37 cm x 34 cm)
Weight 650 lbs (294.8 kg)
Clearances Top 15 in (38.1 cm)
Sides 4 in (10.2 cm)
Rear 6 in (15.2 cm)

Environmental requirements

BTU/Hour output 400 BTU/Hr idle, 1000 BTU/Hr running
Room temperature 68°F to 90°F (20°C to 32°C)
Humidity 10% to 90%, non-condensing
Resources

Outside of Pepper mode, this text will be replaced by the resource cards.

Contact Us

Outside of Pepper mode, this text will be replaced by the content added in the above "Contact a Rep HTML" code editor.

 
Virtual Slide Box Strategic Resource Center

 

Tissue


iView DAB
ultraView