ACCU-CHEK Inform II test strip: results you can trust, delivered in 5 seconds.

New patented technology allows the ACCU-CHEK Inform II test strip and meter to deliver accurate, reliable test results in 5 seconds, with no maltose limitation. The system conducts extensive integrity checks with each test, provides underdose detection and requires a 0.6 μL blood sample.

Test performance that meets today’s requirements for accuracy.

Test results from capillary blood samples using the ACCU-CHEK Inform II system show excellent correlation to reference samples, meeting the ISO accuracy acceptance criteria of ±15 mg/dL for results <75 mg/dL and ±20% for results ≥75 mg/dL.4

         

Principal measurement electrodes

Conduct integrity checks before measuring glucose concentration

Easy-fill tip

Offers capillary end dosing to quickly absorb blood sample

Fill sufficiency electrodes

Ensure enough sample is applied to conduct test and help prevent need for another finger stick

Sensing contacts

The system uses 6 gold electrodes, patented AC/DC technology and a validated algorithm to correct for hematocrit and temperature

Over 150 integrity checks to help ensure accuracy and provide underdose detection.

With each test, the ACCU-CHEK Inform II system runs complete meter, test strip and sample checks in a matter of seconds, to provide underdose detection and help ensure accurate test results. Each time a test strip is inserted, the meter runs initial system and strip checks before giving the signal to apply blood to the strip, helping to prevent the need for an additional finger stick.